Introduction

Affected products

BD MAX System

Reason

BD has confirmed based on customer complaint investigations that a limited number of BD max systems were manufactured with optical filters that have shown a potential to produce a false positive result for Candida glabrata when processing samples positive for Trichomonas vaginalis during testing with the BD max vaginal panel. To prevent any future occurrence BD will be replacing all impacted readers.