Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
151014
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
Biograph mMR System10433372

Introduction

Affected products

Biograph mMR System

Reason

Siemens is informing Biograph mMR users of a potential performance issue with the BodyCOMPASS functionality for pet motion compensation. The issue occurs when BodyCOMPASS is used for elastic motion correction (BodyCOMPASS is an option for syngo E11p). Due to an issue with the resampling of the motion vectors, the BodyCOMPASS does not work for patient orientations other than Head First Supine (HFS). Furthermore, also in HFS patient orientation the motion correction is not working efficiently. As a result the pet motion correction when using BodyCOMPASS delivers suboptimal results.