Health Products

Nara Hospital Bassinet (2018-10-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Stryker Medical
3800 E Centre Avenue
Portage
49002
UNITED STATES
Nara Hospital Bassinet4402-000-002

Introduction

Affected products

Nara Hospital Bassinet

Reason

It was identified [via customer complaint in May 2018] that the volume of air inside the mattress may expand in regions of high altitude which may result in an unintended gap between the plateau of the inflated patient surface and tub.