Health Products

Centricity Universal Viewer (2018-10-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2090255-001
Manufacturer
GE Healthcare
500 W Monroe Street
Chicago
60661
Illinois
UNITED STATES
Centricity Universal Viewer2088034-X

Introduction

Affected products

Centricity Universal Viewer

Reason

An error can occur in Centricity Universal Viewer during study modification (delete/merge/split) activities that prevents synchronization of these updates with the Centricity Enterprise Archive (EA) database. When a diagnostic viewer (e.g. Centricity Universal Viewer Zero Footprint) is connected directly to the Centricity EA and is used for image interpretation, there is the possibility of viewing studies with incorrect patient images because the archive does not contain the updated study.