Health Products

ARCHITECT Total and Free T3 Reagent Kit (2018-10-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Abbott Ireland Diagnostic Division
Lisnamuck, Longford
IRELAND
D. ARCHITECT Total T3 Reagent Kit7K64-27
All lots.
Manufacturer
Abbott Ireland Diagnostic Division
Lisnamuck, Longford
IRELAND
B. ARCHITECT Free T3 Reagent Kit7K63-20
All lots.
Manufacturer
Abbott Ireland Diagnostic Division
Lisnamuck, Longford
IRELAND
C. ARCHITECT Free T3 Reagent Kit7K63-27
All lots.
Manufacturer
Abbott Ireland Diagnostic Division
Lisnamuck, Longford
IRELAND
A. ARCHITECT Total T3 Reagent Kit7K64-20

Introduction

Affected products
  1. ARCHITECT Total T3 Reagent Kit
  2. ARCHITECT Free T3 Reagent Kit
  3. ARCHITECT Free T3 Reagent Kit
  4. ARCHITECT Total T3 Reagent Kit

Reason

During non-conformance investigation a Reagent Cross Contamination (RCC) has been observed for the 6 and 2-point ARCHITECT Free T3 and Total T3 assays on the i1000, when tested alongside assays containing Poly-L-Lysine (PLL) e.g. ARCHITECT TSH. RCC has also been observed for the 2 and 6-point ARCHITECT Free T3 and Total T3 assays when tested on the i2000/i2000SR system and the Alinity i System.