During non-conformance investigation a Reagent Cross Contamination (RCC) has been observed for the 6 and 2-point ARCHITECT Free T3 and Total T3 assays on the i1000, when tested alongside assays containing Poly-L-Lysine (PLL) e.g. ARCHITECT TSH.
RCC has also been observed for the 2 and 6-point ARCHITECT Free T3 and Total T3 assays when tested on the i2000/i2000SR system and the Alinity i System.