Health Products

Compress Device Centering Sleeve, Anti-Rotation Spindle, Anti-Rotation Elliptical Spindle, and Compress Instrument Drill for Anti-Rotation Pin (2018-10-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Biomet Orthopedics
56 East Bell Drive, P.O. BOX 587
Warsaw
46581
UNITED STATES
D. Compress Instrument Drill for Anti-Rotation Pin32-481123
046290
Manufacturer
Biomet Orthopedics
56 East Bell Drive, P.O. BOX 587
Warsaw
46581
UNITED STATES
B. Compress Device Anti-Rotation Spindle178350
282760
Manufacturer
Biomet Orthopedics
56 East Bell Drive, P.O. BOX 587
Warsaw
46581
UNITED STATES
C. Compress Device Anti-Rotation Elliptical Spindle178359
285210
Manufacturer
Biomet Orthopedics
56 East Bell Drive, P.O. BOX 587
Warsaw
46581
UNITED STATES
A. Compress Device Centering Sleeve178537

Introduction

Affected products
  1. Compress Device Centering Sleeve
  2. Compress Device Anti-Rotation Spindle
  3. Compress Device Anti-Rotation Elliptical Spindle
  4. Compress Instrument Drill for Anti-Rotation Pin

Reason

Zimmer Biomet is conducting a medical device correction to update the surgical technique for the compress system. Users should be aware that changes were made to the selection criteria section (pin placement table) on page 42 regarding which array to use in the placement of the anti-rotation pins. The updated pin placement table ensures that the selection criteria for the spindle and centering sleeves will not interfere with the anti-rotation drill or pin. There is no replacement or modification to the product offered.