Dimension Vista System-Lactate Dehydrogenase Flex (LDI) Assay (2018-10-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
Dimension Vista System-Lactate Dehydrogenase Flex (LDI) AssayK2054

Introduction

Affected products

Dimension Vista System-Lactate Dehydrogenase Flex (LDI) Assay

Reason

Siemens Healthcare diagnostics has confirmed that Dimension Vista Lactate Dehydrogenase (LDH) may exhibit an erroneously elevated result when the preceding assay in the reaction cuvette is Lactic Acid (LA). The magnitude of the elevation may be variable. Both quality control and patient samples may be impacted. Incomplete removal of residual lactate dehydrogenase, a component of the la reagent, has been identified as the cause of the potentially elevated results.