Cardiosave Hybrid IABP (2018-09-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
UNITED STATES
Cardiosave Hybrid IABP0998-00-0800-XX

Introduction

Affected products

Cardiosave Hybrid IABP

Reason

Maquet/Getinge has received complaints involving the Cardiosave IABPs regarding the use of certain Intra-Aortic Balloons (IABs) at altitudes above 975 Meters. The Cardiosave may not successfully complete the autofill process required to initiate pumping. This failure may result in either interruption of therapy upon the first maintenance autofill or the inability to start therapy. This condition could potentially lead to patient hemodynamic instability. There have been no adverse events or deaths attributed to this issue. IABs used with the CS100 or CS300IAPB are not affected by the issue.