Health Products

VITROS Chemistry Products 5,1 FS Chemistry System - dLDL Assay (2018-10-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
GEN: 26-6396VITROS Chemistry Products 5,1 FS Chemistry System - dLDL Assay
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
UNITED STATES
680 1728

Introduction

Affected products

VITROS Chemistry Products 5,1 FS Chemistry System - dLDL Assay

Reason

Ortho Clinical Diagnostics (Ortho) has initiated this recall following complaints received for GENs 26 and 27 as well as internal testing of a new lot that determined that GENs 26 and 27 are negatively biased by approximately 15 to 20 mg/dL (0.38 to 0.51 mmol/L) across the measuring range.