Health Products

Artis zee biplane, Artis Q and Artis Q.zen (2018-10-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
AllB. Artis Q
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
10848282
AllC. Artis Q.zen
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
10848355
AllA. Artis zee biplane
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
10094141

Introduction

Affected products
  1. Artis zee biplane
  2. Artis Q
  3. Artis Q.zen

Reason

In affected Artis Systems, vibrations of the C-arm of plane B may result in reduced image quality. As a root cause the ceiling bearing and screw connection to the c-arm in the ceiling support, known as the rotation unit, has been identified. If the connection which consists of twelve screws is not mounted exactly as specified, this may lead to an increase in vibrations which may have a negative impact on image quality. The effects on image quality are not obvious immediately, but will increase over time. This issue can be attributed to one dedicated production batch of ceiling bearings.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime