Revolution CT (2018-10-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
00000440051CN1
Manufacturer
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
Revolution CT5590000

Introduction

Affected products

Revolution CT

Reason

GE Healthcare has become aware of a localization error on revolution CT systems equipped with the smartstep option. The z location displayed on image annotation and on the gantry display may not be the actual location of the table. This can occur if:
  1. The table has been moved all the way out of the gantry using the 'out' button on the gantry control or the 'unload pedal'.
  2. The table is then moved back into the gantry using the 'move to start location' button on the hand held controller.


In this situation, if a scan is initiated by the user, the scan may be acquired in the wrong location. An additional scan may be necessary.