VITROS 5, 1 FS Chemistry System, VITROS 3600 Immunodiagnostic System, VITROS 5600 Integrated System, and VITROS 4600 Chemistry System (2018-10-21) Oct 21, 2018 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Brand name Companies Model or catalog number More than 10 numbers, contact manufacturer. D. VITROS 5600 Integrated System Manufacturer Ortho Clinical Diagnostics Inc. 100 Indigo Creek Drive Rochester 14626 New York UNITED STATES 680 2413 More than 10 numbers, contact manufacturer. E. VITROS 4600 Chemistry System Manufacturer Ortho Clinical Diagnostics Inc. 100 Indigo Creek Drive Rochester 14626 New York UNITED STATES 680 2445 More than 10 numbers, contact manufacturer. B. VITROS 5,1 FS Chemistry System (Refurbished) Manufacturer Ortho Clinical Diagnostics Inc. 100 Indigo Creek Drive Rochester 14626 New York UNITED STATES 6801890 J36000773 C. VITROS 3600 Immunodiagnostic System Manufacturer Ortho Clinical Diagnostics Inc. 100 Indigo Creek Drive Rochester 14626 New York UNITED STATES 680 2783 More than 10 numbers, contact manufacturer. VITROS 5,1 FS Chemistry System Manufacturer Ortho Clinical Diagnostics Inc. 100 Indigo Creek Drive Rochester 14626 New York UNITED STATES 680 1375
Affected products
VITROS 5,1 FS Chemistry System
VITROS 5,1 FS Chemistry System (Refurbished)
VITROS 3600 Immunodiagnostic System
VITROS 5600 Integrated System
VITROS 4600 Chemistry System
Under specific scenarios, a software anomaly will occur causing the VITROS Systems to incorrectly determine the reagent's on-analyzer stability time. When the anomaly occurs, the system could use reagents past their on-analyzer stability time without alerting the user with a condition code or flagging the associated results with an RE (Reagent Expired) code. In the event reagents are used beyond their specific on-analyzer stability times, the associated test results may be affected.