CDI Blood Parameter Monitoring System 500 H/Sat Probe Assembly (2018-10-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
500AHCT
Manufacturer
Terumo Cardiovascular Systems Corp.
6200 Jackson Road
Ann Arbor
48103
Michigan
UNITED STATES
CDI Blood Parameter Monitoring System 500 H/Sat Probe Assembly500A HCT

Introduction

Affected products

CDI Blood Parameter Monitoring System 500 H/Sat Probe Assembly

Reason

A specific subset of the h/sat probe assembly population is being withdrawn due to a supplied component in the optics, the optical module. A change by the supplier to a different material and anodization resulted in increased failures. The CDI System 500 monitor conducts an automated self-test when the device is powered on. If the monitor contains an affected component, a color chip test failure will occur and the h/sat probe will be disabled.