Kodak DirectView DR 7500 System (2018-10-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
S/N: 150
Manufacturer
Carestream Health, Inc.
150 Verona Street
Rochester
14608
New York
UNITED STATES
Kodak DirectView DR 7500 System879 1345

Introduction

Affected products

Kodak DirectView DR 7500 System

Reason

SEDECAL S.A. notified Carestream Health Inc. of a potential safety issue that could lead to the breakage of the two steel cables that attach the equipment to its ceiling anchor. The main cause of this problem is the excessive wear of the vertical pulley where the steel cables are located. When the wear is not uniform in both grooves of the pulley, the distance between the steel cables is not constant. In the most unfavorable case, this irregular wear could cause the cables to rub against each other over time to the point of rupture.