SEDECAL S.A. notified Carestream Health Inc. of a potential safety issue that could lead to the breakage of the two steel cables that attach the equipment to its ceiling anchor. The main cause of this problem is the excessive wear of the vertical pulley where the steel cables are located. When the wear is not uniform in both grooves of the pulley, the distance between the steel cables is not constant. In the most unfavorable case, this irregular wear could cause the cables to rub against each other over time to the point of rupture.