Health Products

Stockert Heater Cooler System 3T (2018-10-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer
Manufacturer
Livanova Deutschland GmbH
Lindberghstrasse 25
Munich
80939
GERMANY
Stockert Heater Cooler System 3T16-02-81

Introduction

Affected products

Stockert Heater Cooler System 3T

Reason

LivaNova is providing updated instructions to monitor the concentration of hydrogen peroxide in the water circuit of the 1T and 3T Heater-Cooler devices, to verify that sufficient concentration of hydrogen peroxide is present to limit microbial growth, and to adjust the concentration of hydrogen peroxide if it drops below 100 ppm. This regimen enhances the water maintenance regimen outlined in Section 6.4 of the operating instructions.