VITROS ECi Immunodiagnostic Products (2018-11-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
ALL EXPIRED/IN-DATE/FUTURE LOT
Manufacturer
Ortho Clinical Diagnostics
Felindre Meadows
Pencoed, Bridgend
CF35 5PZ
UNITED KINGDOM
F. VITROS ECi Immunodiagnostic Products Free PSA Reagent Pack684 2845
ALL EXPIRED/IN-DATE/FUTURE LOT
Manufacturer
Ortho Clinical Diagnostics
Felindre Meadows
Pencoed, Bridgend
CF35 5PZ
UNITED KINGDOM
D. VITROS ECi Immunodiagnostic Products FSH Reagent Pack193 1922
ALL EXPIRED/IN-DATE/FUTURE LOT
Manufacturer
Ortho Clinical Diagnostics
Felindre Meadows
Pencoed, Bridgend
CF35 5PZ
UNITED KINGDOM
E. VITROS ECi Immunodiagnostic Products Folate Reagent Controls & Calibrators151 3266
ALL EXPIRED/IN-DATE/FUTURE LOT
Manufacturer
Ortho Clinical Diagnostics
Felindre Meadows
Pencoed, Bridgend
CF35 5PZ
UNITED KINGDOM
B. VITROS ECi Immunodiagnostic Products TSH Reagent Pack191 2997
ALL EXPIRED/IN-DATE/FUTURE LOT
Manufacturer
Ortho Clinical Diagnostics
Felindre Meadows
Pencoed, Bridgend
CF35 5PZ
UNITED KINGDOM
C. VITROS ECi Immunodiagnostic Products Prolactin Reagent Pack184 9793
ALL EXPIRED/IN-DATE/FUTURE LOT
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Felindre Meadows, Pencoed, Bridgend
Great Britain
CF35 5PZ
UNITED KINGDOM
A. VITROS ECi Immunodiagnostic Products NTproBNP Reagent Pack680 2156

Introduction

Affected products
  1. VITROS ECi Immunodiagnostic Products NTproBNP Reagent Pack
  2. VITROS ECi Immunodiagnostic Products TSH Reagent Pack
  3. VITROS ECi Immunodiagnostic Products Prolactin Reagent Pack
  4. VITROS ECi Immunodiagnostic Products FSH Reagent Pack
  5. VITROS ECi Immunodiagnostic Products Folate Reagent Controls & Calibrators
  6. VITROS ECi Immunodiagnostic Products Free PSA Reagent Pack

Reason

Ortho Clinical Diagnostics determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions for Use (IFU).

Historically, these assays were evaluated for the interference consistent with CLSI document EP7. Biotin was tested and was not found to cause a bias of >/=10%.

To assess the potential interference of biotin supplements on VITROS MicroWell Assays, Ortho performed more robust testing using the new CLSI guidelines EP7 and EP37 and determined that the information in the current IFUs is no longer supported for these five (5) assays. The concentration of biotin that causes >/=10% bias is listed in the customer letter.