Health Products

Access hsTnI Reagent (2018-10-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Immunotech SAS
130 AVENUE DE LATTRE DE TASSIGNY
MARSEILLE CEDEX 09,
13276
FRANCE
Access hsTnI ReagentB52699

Introduction

Affected products

Access hsTnI Reagent

Reason

The Cardiac Troponin T Cross Reactivity concentration of 253 ng/ml (253,000 pg/ml) previously listed in the access HSTNI Instructions For Use (IFU) does not apply to samples with HSTNI concentrations less than 16 pg/ml (ng/l).

Samples containing cardiac troponin t concentrations > 125 ng/ml (125,000 pg/ml) may cross react with the access HSTNI assay:
  • access HSTNI results may change by more than 10% for HSTNI sample concentrations greater than 11.5 pg/ml (ng/l);
  • access HSTNI results may change by more than 2.30 pg/ml (ng/l) for HSTNI sample concentrations less than or equal to 11.5 pg/ml (ng/l).