Health Products

DELFIA Xpress  (2018-11-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
60000654
Manufacturer
Wallac Oy
Mustionkatu 6, P.O. Box 10
Turku
20750
FINLAND
DELFIA Xpress6000-0010

Introduction

Affected products

DELFIA Xpress

Reason

The bushing components of the sample tube racks, delivered with the instrument, are deficient; the plastic compound of the bushing component omitted a required ingredient. The deficiency causes tilting of the sample tubes.