Affected products
A. PreciseType HEA Molecular BeadChip Test
B. RHD Molecular BeadChip Test
As a result of a complaint investigation, we have determined that the performance of the HEA PreciseType Beadchip Kit (IVD) listed above may be impacted by sporadic, decreased activity of the Clean-up Reagent, PN 800-00191, Lot 18-319A contained within these kits. Customers may observe an increase in occurrence of IC/LS calls when using these kits. In some instances, an incorrect call may be assigned. Additionally, although there has been no observed impact to the RHD BeadChip Kit PN 800-10220-48, Lot 19-28-C, Immucor has decided to discontinue use of Clean-up Reagent, PN 800-00191, Lot 18-319A in all BeadChip products and provide replacement Clean-up reagent to customers who have BeadChip kit lots that contain this lot of Clean-up reagent.