Health Products

RadPRO Mobile 3 40kW (2018-11-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
G65175
Manufacturer
Virtual Imaging, Inc.
720 SOUTH POWERLINE ROAD, SUITE E
DEERFIELD BEACH
33442
Florida
UNITED STATES
RadPRO Mobile 3 40kWMOBILE 3 40KW 50G/C

Introduction

Affected products

RadPRO Mobile 3 40kW

Reason

There is the potential for the four (4) retention screw brass tabs, which secure the system's collimator to the tube head mount, to wear out which could result in the system's collimator separating from the tube head mount and potentially falling off of the system.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime