Electronic O2 Blender/Analyzer (2018-11-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
contact manufacturer
Manufacturer
Terumo Cardiovascular Systems Corp.
6200 Jackson Road
Ann Arbor
48103
Michigan
UNITED STATES
Electronic O2 Blender/Analyzer801188

Introduction

Affected products

Electronic O2 Blender/Analyzer

Reason

There is an intermittent failure that occurred with certain devices assembled with stepper motors that did not meet supplier specification. This causes a blender initialization fault upon start-up, or during an EPGS calibration attempt. System diagnostics recognizes this and posts an error message 'Gas System Knob Adjust Only' at which time the gas system flow and FiO2 control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs.