Health Products

MagNA Pure 24 Instrument (2018-10-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Serial no. 1016MagNA Pure 24 Instrument
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
07290519001

Introduction

Affected products

MagNA Pure 24 Instrument

Reason

There is a potential for cross-contamination when using specific protocols (HGDNA 200 version 1, HGDNA DS 200 version 1, fast HGDNA 200 version 1 and HGDNA 1000 protocols version 2) on the MagNA Pure 24 System.  A new version of the protocol set (version 1.4) is available. It includes updated versions of the HGDNA 200 (version 2.0), HGDNA ds 200 (version 2.0), fast HGDNA 200 (version 2.0) and HGDNA 1000 protocols (version 3.0).   The protocols have been updated to exclude the risk of a potential cross-contamination without affecting the purification performance of the HGDNA protocols.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime