Health Products

Xper Flex Cardio Physiomonitoring System-Main Unit (2018-11-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Invivo, a Div. of Philips Medical Systems
12151 Research Parkway
Orlando
32826
Florida
UNITED STATES
Xper Flex Cardio Physiomonitoring System-Main UnitFC2010

Introduction

Affected products

Xper Flex Cardio Physiomonitoring System-Main Unit

Reason

Philips has become aware of the following issues:
1. Real time ECG and pressure waveform anomaly: under certain conditions, real time waveforms may not be accurately plotted or displayed. This problem is related to the display of waveforms only, and all numeric data including Invasive Blood Pressure (IBP), NIBP, Respiratory Rate, and End Tidal CO2 are correct. Should this occur during the plotting of an ECG QRS complex on the ECG channel, the r wave may be shortened or missing or the QRS complex may appear slightly broader. A pressure or other phasic waveform may appear to have a "notch" or a "flattened" appearance.
2. SPO2 measurement may not update on the display: the oxygen saturation (SPO2) numeric value displayed on the device may freeze, resulting in the display of measurements that are not current.
3. NIBP measurement may not update on the display: in auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the Non-Invasive Blood Pressure (NIBP) numeric value displayed on the device would not update, resulting in the display of measurements that are not current.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime