Health Products

SQ-RX Pulse Generator (2018-11-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Boston Scientific Corporation
4100 Hamline Avenue North
St. Paul, Minnesota
55112-5798
Minnesota
UNITED STATES
SQ-RX Pulse Generator1010

Introduction

Affected products

SQ-RX Pulse Generator

Reason

Potential for a shortened replacement interval after a Charge Time (CT) / Battery Depletion (BD) alert has occurred or after the battery status reaches Elective Replacement Indicator (ERI) in the first-generation Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) system's SQ-Rx model 1010 Pulse Generator (PG). Boston scientific have observed an elevated rate of these PGs experiencing latent battery malfunctions resulting in accelerated battery depletion and a shortened replacement interval.