Potential for a shortened replacement interval after a Charge Time (CT) / Battery Depletion (BD) alert has occurred or after the battery status reaches Elective Replacement Indicator (ERI) in the first-generation Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) system's SQ-Rx model 1010 Pulse Generator (PG). Boston scientific have observed an elevated rate of these PGs experiencing latent battery malfunctions resulting in accelerated battery depletion and a shortened replacement interval.