Affected products
Advanced Perfusion System 1 - EPGS Service Manual
This notice is to clarify the Terumo Advanced Perfusion System 1 service manual for the Electronic Patient Gas System (EPGS), catalog # 875306. The service manual (rev. E) states that the oxygen sensor expiration alert is triggered at 300,000 %-hours, whereas the EPGS software (version 1.30) presents an alert message and prevents calibration of the EPGS when the oxygen sensor has reached 100,000 %-hours, which accurately reflects the oxygen sensor design specification. The Terumo System 1 Service Manual (EPGS) discrepancy is limited to service documentation only.