Signa Horizon LX Magnetic Resonance System, Signa Infinity 1.5T TwinSpeed Excite MR System, Signa HDxt 1.5T and 3.0T MR System (2018-11-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
00000007532YR5D. Signa HDxt 3.0T MR System
Manufacturer
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
SIGNA HDXT 3.0T MR SYSTEM
9520M28B. Signa Infinity 1.5T TwinSpeed Excite MR System
Manufacturer
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
M3000TD
00000283347MR3C. Signa HDxt 1.5T MR System
Manufacturer
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
1.5T SIGNA HDXT MR SYSTEM
0000098-1219-2A. Signa Horizon LX Magnetic Resonance System
Manufacturer
GE Medical Systems LLC
3000 North Grandview Blvd
Waukesha
53188
Wisconsin
UNITED STATES
46-265300G1

Introduction

Affected products

A. Signa Horizon LX Magnetic Resonance System
B. Signa Infinity 1.5T TwinSpeed Excite MR System
C. Signa HDxt 1.5T MR System
D. Signa HDxt 3.0T MR System

Reason

Latest on-site software version may not have been reinstalled at some sites after service activities were performed that required software to be reloaded.
If the incorrect software version is loaded on the Brivo MR355, Optima 360, Signa Profile, Signa Ovation HD, Signa Ovation .35T, Signa HFO, Signa HDE, Signa Ovation with Excite, 0.7T Signa Openspeed, Signa Infinity with Excite Technology, Signa Excite 3T and 1.5T Signa HDXT Systems the images can be flipped left/right and/or there can be patient data mismatch.
If the incorrect version is loaded on the Signa Infinity or the Signa Infinity Twinspeed System this could result in slice mis-registration.
If the incorrect software version is loaded on the Signa Creator or the Signa Explorer, and the user initiates a TPS (transceiver processing and storage) reset during a patient scan, the RF power monitor will be disabled for the remainder of the scan and unable to detect a subsequent failure of the RF transmit function. This could result in the in higher than expected thermal dose and localized heating to the patient.
If the incorrect software version is loaded on the GE Signa OpenSpeed System a patient's hand may be pinched if the patients puts his/her hand between the on top of the coil (CTL array and body flex) and ceiling of the magnet bore while the patient table is moving into the bore.