Health Products

T34 Ambulatory Syringe Pump (2018-11-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device, Biologic/vaccine
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.T34 Ambulatory Syringe Pump
Manufacturer
ENDYMED MEDICAL LTD.
12 LESHEM STREET
CAESAREA
3088900
ISRAEL
T34

Introduction

Affected products

T34 Ambulatory Syringe Pump

Reason

It has been identified that there can be a 2mm +/- overall length difference between various manufacturers' batteries that can be used in CME's T34TM Ambulatory Syringe Pumps. This could result in movement within the battery housing leading to a possible loss of connection, resulting in the pump powering down and a possible under infusion of pain medication occurring.