Health Products

Reliance Blunt Probe (2018-11-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
17E168Reliance Blunt Probe
Manufacturer
Stryker Spine
2 Pearl Court
Allendale, New Jersey
07401
New Jersey
UNITED STATES
48066103

Introduction

Affected products

Reliance Blunt Probe

Reason

During a Quality Control inspection of the spherical diameter on the Reliance Blunt Probe, it was detected that parts were not conforming. Specifically, instead of having a spherical diameter, the probe was observed to have 2 radii and a flat surface. Additionally, it was observed that some parts exhibited an edge on the intersection between the radius and the diameter of the probe.