Affected products
A. Hydra Vision Urology System
B. Direct Digital Imaging System
Sedecal, the generator manufacturer, created a software algorithm to estimate and display the approximate dose of radiation delivered. Since the device is designed to deliver 50% less radiation when placed in child/pediatric mode, a separate algorithm had to be developed to display this lesser amount. Although the generator's radiation output in all modes was as intended and within specification, when placed in child/pediatric auto-exposure mode, the software selected the adult display algorithm instead of the child/pediatric display algorithm, resulting in a console display reading that was higher than what was actually delivered.