CARESCAPE Central Station (2018-01-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
GE Medical Systems Information Technologies, Inc.
8200 W Tower Ave
Milwaukee
53223
Wisconsin
UNITED STATES
CARESCAPE Central Station2051117-005

Introduction

Affected prodcuts

CARESCAPE Central Station

Reason

GE Healthcare has recently become aware that after boot-up or system start up, CSCS v2.0.2 units may display unexpected no comm notifications in the patient multi-viewer or single viewer window when the CSCS is connected to both the mission critical (mc) and the information exchange (ix) networks. If a no comm notification is displayed, all patient waveforms, parameter data and alarms are lost at the CSCS and patients admitted to telemetry transmitters are unmonitored unless monitored via other central stations, bedside monitors or direct clinical observation.