Health Products

Compress Device Centering Sleeve, Compress Device and Compress Device Short Anti-Rotation Spindle, Anti-Rotation Elliptical Spindle, and Compress Instrument Drill for Anti-Rotation Pin (2018-12-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
AllF. Compress Instrument Drill for Anti-Rotation Pin
Manufacturer
Biomet Orthopedics
56 East Bell Drive, P.O. BOX 587
Warsaw
46581
UNITED STATES
32-481123
115070D. Compress Device Short Anti-Rotation Spindle
Manufacturer
Biomet Orthopedics
56 East Bell Drive, P.O. BOX 587
Warsaw
46581
UNITED STATES
178362
AllE. Compress Device Short Anti-Rotation Elliptical Spindle
Manufacturer
Biomet Orthopedics
56 East Bell Drive, P.O. BOX 587
Warsaw
46581
UNITED STATES
178371
046290B. Compress Device Anti-Rotation Spindle
Manufacturer
Biomet Orthopedics
56 East Bell Drive, P.O. BOX 587
Warsaw
46581
UNITED STATES
178351
282760C. Compress Device Anti-Rotation Elliptical Spindle
Manufacturer
Biomet Orthopedics
56 East Bell Drive, P.O. BOX 587
Warsaw
46581
UNITED STATES
178360
More than 10 numbers, contact manufacturer.A. Compress Device Centering Sleeve
Manufacturer
Biomet Orthopedics
56 East Bell Drive, P.O. BOX 587
Warsaw
46581
UNITED STATES
178537

Introduction

Affected products
  1. Compress Device Centering Sleeve
  2. Compress Device Anti-Rotation Spindle
  3. Compress Device Anti-Rotation Elliptical Spindle
  4. Compress Device Short Anti-Rotation Spindle
  5. Compress Device Short Anti-Rotation Elliptical Spindle
  6. Compress Instrument Drill for Anti-Rotation Pin

Reason

Zimmer Biomet is expanding a previous medical device notification, dated October 22, 2018, in order to update the pin placement tables on page 42 and page 45 of the compress and mini compress surgical techniques, respectively, regarding which array to use in the placement of the anti-rotation pins. These updates were made due to the potential for interference between the centering sleeves and anti-rotation drill when using certain spindle and centering sleeve combinations. Additional item numbers are included in the updated notice to users.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime