Terumo Advanced Perfusion System 1 Power Manager Board
801763
Introduction
Affected products
Terumo Advanced Perfusion System 1 Power Manager Board
Reason
During internal testing, Terumo CVS identified that screws on a Terumo Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which is not properly torqued can loosen over time resulting in an electrical short.