Ventri Nuclear Medicine System and Discovery NM 530c Nuclear Medicine System (2018-12-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device, Biologic/vaccine
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.B. Discovery NM 530c Nuclear Medicine System
Manufacturer
GE Medical Systems Israel
Functional Imaging
4 Hayozma Street
S8000RA
More than 10 numbers, contact manufacturer.A. Ventri Nuclear Medicine System
Manufacturer
GE Medical Systems Israel
Functional Imaging
4 Hayozma Street
H3000YT

Introduction

Affected products
  1. Ventri Nuclear Medicine System
  2. Discovery NM 530c Nuclear Medicine System

Reason

GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient's finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.

The relevant software update, Field Action 40847, released in 2013, issued a mandatory software update to disable the automatic Unload motion feature. Requiring Unload table motions to be done manually (via Remote Control Unit) necessitates that the operator remain in close proximity to the patient during the Unload process. Appropriate operator monitoring while unloading the patient helps ensure that the patient arms and hands do not protrude from the table.