GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient's finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.
The relevant software update, Field Action 40847, released in 2013, issued a mandatory software update to disable the automatic Unload motion feature. Requiring Unload table motions to be done manually (via Remote Control Unit) necessitates that the operator remain in close proximity to the patient during the Unload process. Appropriate operator monitoring while unloading the patient helps ensure that the patient arms and hands do not protrude from the table.