Health Products

Ingenia Elition X and S (2018-12-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A. Ingenia Elition X781357
Not applicable
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684 PC
NETHERLANDS
A. Ingenia Elition X781358

Introduction

Affected products
  1. Ingenia Elition X
  2. Ingenia Elition S

Reason

Philips has become aware of a potential hazardous situation with the VitalScreen 'Stop Table' software button on the affected products (the yellow button on the right bottom corner). During automatic tabletop movement, pressing and holding the 'Stop Table' button may lead to an unexpected restart of the table movement.