ADVIA Centaur and ADVIA Centaur CP (2018-12-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Lot 04642306
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. ADVIA Centaur CP10492399
Lot 04642306
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. ADVIA Centaur10492399

Introduction

Affected products
  1. ADVIA Centaur
  2. ADVIA Centaur CP

Reason

Siemens identified a positive bias with ADVIA Centaur aTG Kit lots ending in 316 and lower when compared to the standardization to world health organization (who) reference preparation MRC 65/93 stated in the Instructions For Use (IFU).

Traceability to who reference preparation MRC 65/93 is restored with the release of ADVIA Centaur aTG kit lots ending in 318 and higher (available in December 2018). As stated in the IFU, the theoretical who international units (IU/ml) is on average 3-fold higher than Siemens healthcare diagnostics standardization. Moving forward, this traceability will be maintained through enhancements to the control system.

Customers will observe a negative shift in patient results when transitioning from ADVIA Centaur aTG reagent kit lots 316 and lower to ADVIA centaur aTG reagent kit lots 318 and higher.