Siemens identified a positive bias with ADVIA Centaur aTG Kit lots ending in 316 and lower when compared to the standardization to world health organization (who) reference preparation MRC 65/93 stated in the Instructions For Use (IFU).
Traceability to who reference preparation MRC 65/93 is restored with the release of ADVIA Centaur aTG kit lots ending in 318 and higher (available in December 2018). As stated in the IFU, the theoretical who international units (IU/ml) is on average 3-fold higher than Siemens healthcare diagnostics standardization. Moving forward, this traceability will be maintained through enhancements to the control system.
Customers will observe a negative shift in patient results when transitioning from ADVIA Centaur aTG reagent kit lots 316 and lower to ADVIA centaur aTG reagent kit lots 318 and higher.