Health Products

MAGNETOM Prisma and Prisma Fit (2018-12-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
166026
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
B. MAGNETOM Prisma10849583
167006
Manufacturer
Siemens Healthcare GmbH
Henkestr. 127
Erlangen
91052
GERMANY
A. MAGNETOM Prisma Fit10849583

Introduction

Affected products
  1. MAGNETOM Prisma Fit
  2. MAGNETOM Prisma

Reason

Siemens has identified a software update in order to fully satisfy the IEC 60601-2- 33 standard. This standard defines a limit to avoid cardiac stimulations incorporating a well-spaced safety factor. With the currently installed MAGNETOM Prisma software, this given limit may be exceeded in rare cases.