Health Products

Freedom EVO 75, 100, 150, and 200 (2018-11-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Tecan Schweiz AG
Seestrasse 103
Mannedorf
8708
SWITZERLAND
D. Freedom EVO 15010641150
More than 10 numbers, contact manufacturer.
Manufacturer
Tecan Schweiz AG
Seestrasse 103
Mannedorf
8708
SWITZERLAND
B. Freedom EVO 7520738668
More than 10 numbers, contact manufacturer.
Manufacturer
Tecan Schweiz AG
Seestrasse 103
Mannedorf
8708
SWITZERLAND
C. Freedom EVO 10010641100
More than 10 numbers, contact manufacturer.
Manufacturer
Tecan Schweiz AG
Seestrasse 103
Mannedorf
8708
SWITZERLAND
A. Freedom EVO 20010641200

Introduction

Affected products

A. Freedom EVO 200
B. Freedom EVO 75
C. Freedom EVO 100
D. Freedom EVO 150

Reason

The affected versions are freedom EVOware 2.5 - 2.7 SP3 under Windows XP and Windows 7, and 2.8 under Windows 7 and Windows l0. Both freedom EVOware standard and EVOware plus are affected, in both limited and extended support configurations. This problem does not exist in earlier or later versions of EVOware.

Starting January 1st 2020 at 00:00 hrs, freedom EVOware will abort a run if any processing errors occur or when freedom EVOware is started in conjunction with touchtools.
  • In case of using touchtools: the system is not able to start a run
  • In case of using EVOware: the system will abort a run when an error message appears and not restart both cases lead to a loss of instrument functionality. There is no risk of safety to operators or of incorrect processing of samples as the script or process stops with error messages.