Sequestra 1000 System (2018-11-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.Sequestra 1000 System
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
BT920

Introduction

Affected products

Sequestra 1000 System

Reason

Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit. There were no (0) adverse reactions or harms associated with these complaints. Medtronic does not recommend any further actions for patients already treated with the affected devices. These patients should continue to be monitored in accordance with standard practice.