Health Products

DRAEGERWERK AG & CO. KGAA's Disposable Breathing Circuits (2018-12-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
H. VentStar Anesthesia Watertrap 180 W/O LuerlockMP00371
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
I. VentStar Anesthesia Watertrap 180MP00372
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
F. VentStar Watertrap (Pediatric) 180 W/O LuerlockMP00362
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
G. VentStar Basic(P)250MP00352
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
D. VentStar Basic 250MP00350
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
E. VentStar Basic (P)180MP00351
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
B. VentStar Watertrap 180 W/O LuerlockMP00334
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
K. VentStar Anesthesia Basic 180MP00365
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
C. VentStar Basic 180MP00349
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
L. VentStar Anesthesia Basic 250MP00366
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
A. VentStar Watertrap 180MP00337
More than 100 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO. KGAA
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
J. VentStar Anesthesia Watertrap (P) 180MP00374

Introduction

Affected products
  1. VentStar Watertrap 180
  2. VentStar Watertrap 180 W/O Luerlock 
  3. VentStar Basic 180
  4. VentStar Basic 250
  5. VentStar Basic (P)180
  6. VentStar Watertrap (Pediatric) 180 W/O Luerlock 
  7. VentStar Basic(P)250
  8. VentStar Anesthesia Watertrap 180 W/O Luerlock 
  9. VentStar Anesthesia Watertrap 180
  10. VentStar Anesthesia Watertrap (P) 180
  11. VentStar Anesthesia Basic 180
  12. VentStar Anesthesia Basic 250

Reason

Two cases have been reported in which patients were not connected to the anesthesia machine due to the incorrect connection of the breathing circuit. The connections of the anesthesia machine and of the Y piece were short- circuited. Consequently, the patients did not receive the expected ventilation; in one of the two cases the patient in question suffered irreversible harm, in the other case the patient experienced a temporary and reversible deterioration in condition.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime