Health Products

Elecsys Anti-TSHR (2018-08-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
28356401
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
Elecsys Anti-TSHR07026951190

Introduction

Affected products

Elecsys Anti-TSHR

Reason

Following internal investigation of customer complaints, the manufacturer confirmed a limitation in On-Board Stability (OBS) of the product Elecsys Anti-TSHR (#07026951190) lot #28356401 used on cobas e 801 module. The OBS could be maintained for 8 weeks, in comparison to the 16 weeks OBS claim as described in the Method Sheet of the assay.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime