Siemens healthcare diagnostics has confirmed that the APTT ratio of the Dade Ci-Trol Heparin Control Low lot 548455 may recover below the expected range for the APTT ratio given within the instructions for use (ratio for Sysmex CA-1500 systems: 1.4 - 2.2). Internal investigation with this specific Dade Ci-Trol Heparin Control Low lot confirmed that the content of unfractionated heparin is very low. This may lead to a low recovery of the APTT ratio with most APTT reagents and also to non-detectable heparin neutralization after treatment with Dade Hepzyme. This renders this control lot unsuitable for some of the intended uses given in the instruction for use.