Health Products

Ci-Tro Heparin Controls Low and High (2018-12-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
548455
Manufacturer
Siemens Healthcare Diagnostics Products Gmbh
Emil-Von-Behring-Str. 76
Marburg
35041
GERMANY
Ci-Tro Heparin Controls Low and HighB4224-50

Introduction

Affected products

Ci-Tro Heparin Controls Low and High

Reason

Siemens healthcare diagnostics has confirmed that the APTT ratio of the Dade Ci-Trol Heparin Control Low lot 548455 may recover below the expected range for the APTT ratio given within the instructions for use (ratio for Sysmex CA-1500 systems: 1.4 - 2.2). Internal investigation with this specific Dade Ci-Trol Heparin Control Low lot confirmed that the content of unfractionated heparin is very low. This may lead to a low recovery of the APTT ratio with most APTT reagents and also to non-detectable heparin neutralization after treatment with Dade Hepzyme. This renders this control lot unsuitable for some of the intended uses given in the instruction for use.