Neurosign V4 Intraoperative Nerve Monitor (2018-12-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
0037
Manufacturer
The Magstim Company Limited
Spring Gardens, Carmarthenshire
Whitland, Wales
SA34 0HR
UNITED KINGDOM
Neurosign V4 Intraoperative Nerve Monitor4230-00

Introduction

Affected products

Neurosign V4 Intraoperative Nerve Monitor

Reason

Customer reported that the device would not function - error message appeared "fatal error occurred". Investigation into the 'fatal error' system error (as shown below) has identified that two brand of SD cards are more vulnerable to this intermittent fault. It has also been confirmed that the system error is caused when insufficient time is provided when switching off the NeuroDign V4.