Alinity i Processing Module (2018-12-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Ai01457
Manufacturer
Abbott GmbH and Co. K.G.
Max Planck Ring 2
Wiesbaden
65205
GERMANY
Alinity i Processing Module03R65-01

Introduction

Affected products

Alinity i Processing Module

Reason

Abbott has identified an issue with the Alinity I gear pump assembly (part number a 30108552-01) resulting in foaming/bubbling out of the bottle reservoir for concentrated wash buffer and an unexpected amount of dried residue of buffer. The foaming/bubbling is a result of an improper calibration of the pump's operating speed.