Health Products

LOGIQ P6 Ultrasound System - Main Unit and LOGIQ P6 Ultrasound System - Console Upgrade to BT09 (2018-12-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.B. LOGIQ P6 Ultrasound System - Console Upgrade to BT09
Manufacturer
GE Ultrasound Korea Ltd.
9 Sunhwan-ro 214beon-gil
Jungwon-gu. Seongnam, Gyeonggi
462-807
SOUTH KOREA
H46172LJ
More than 10 numbers, contact manufacturer.A. LOGIQ P6 Ultrasound System - Main Unit
Manufacturer
GE Ultrasound Korea Ltd.
9 Sunhwan-ro 214beon-gil
Jungwon-gu. Seongnam, Gyeonggi
462-807
SOUTH KOREA
H46082LW

Introduction

Affected products
  1. LOGIQ P6 Ultrasound System - Main Unit
  2. LOGIQ P6 Ultrasound System - Console Upgrade to BT09

Reason

In rare instances, if a certain component fails, there is a potential of increased temperature in the probe-head surface when using these scanners with specific probes. This could result in a burn to the patient in specific types of scans.