Health Products

StealthAir Percutaneous Pin Adapter (2018-12-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Medtronic Navigation Inc.
826 Coal Creek Circle
Louisville
80027
Colorado
UNITED STATES
StealthAir Percutaneous Pin Adapter9734752

Introduction

Affected products

StealthAir Percutaneous Pin Adapter

Reason

Medtronic has become aware that, under certain circumstances, the percutaneous pin adapter used with the StealthAir frame assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and secure. This can occur due to the shallow angle of the percutaneous pin adapter. Frame rotation may cause inaccurate navigation, which can potentially result in misplaced screws or implants, surgical delay, aborted navigation, or patient injury.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime