Leica M220 F12 Microscope System (2018-12-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
170718002
Manufacturer
Leica Microsystems (Schweiz) Ag
Max Schmidheiny-Strasse 201
Heerbrugg
9435
SWITZERLAND
Leica  M220 F12 Microscope System10448367

Introduction

Affected products

Leica  M220 F12 Microscope System

Reason

Leica Microsystems has received two (2) (Nov. 2018) complaints pertaining to the M220 F12 Surgical Microscope System. In both, the optics carrier were reported to have fallen. These complaints detailed mechanical part failures with the screw that connects the optics carrier to the free end of the swing arm. Leica identified the cause of the observed issues to be insufficient securing of the screws that connect the swing arm with the optics carrier.