Health Products

Arrow Pressure Injectable Multi Lumen CVC Kit (2019-01-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
13F17J0444
Manufacturer
Arrow International, Inc. (Subsidiary of Teleflex, Incorporated)
2400 Bernville Road
Reading
19605
UNITED STATES
Arrow Pressure Injectable Multi Lumen CVC KitASK-42703-PRJ

Introduction

Affected products

Arrow Pressure Injectable Multi Lumen CVC Kit

Reason

Arrow International is recalling the product due to a labeling error.

The lot (13F17J0444) is product code ASK-42703-PRJ so it should contain a 3-lumen, 7 Fr x 16 cm catheter. However, the product may be incorrectly labeled as ASK-45703-PRJ.  In this case, the pre-printed lidstock states it contains a 3-lumen, 7 Fr x 20 cm catheter. There is an additional label on the lidstock outlining the correct part number ASK-42703-PRJ. The package contains a 3-lumen, 7 Fr x 16 cm catheter.

The incorrect catheter length would not likely cause any serious health consequence; however, it may lead to a minor delay in completion of the procedure. No patient injuries have been reported related to this issue.