Philips has received a report in which the lithium ion battery in a Philips PageWriter TC cardiograph overheated and ignited. In this case, the battery had exceeded its life expectancy and should have been replaced when the number of charge-discharge cycles first exceeded 300 cycles or when the remaining battery capacity fell below 80% of that of a new battery. Although the Philips PageWriter TC cardiographs can display actual information on the battery State of Health (SOH) and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery.