Amia Automated PD System - Cycler (2018-07-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer
Manufacturer
Baxter Healthcare Corporation
1 Baxter Parkway
Deerfield, Illinois
60015
UNITED STATES
Amia Automated PD System - Cycler5C9320

Introduction

Affected Products

Amia Automated PD System - Cycler

Reason

Baxter Corporation has identified a potential failure with the Amia cycler where the power cord socket can become loose or detached when the power cord is unplugged from the back of the cycler. To address this issue, Baxter will be inspecting and repairing any loose power cord sockets during the next Amia cycler service. The Amia cyclers may continue to be used, and do not need to be returned at this time.