Health Products

Medtronic Adapta Pacing System (Adapta, Sensia, Attesta) (2019-01-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis, Minnesota
55432
UNITED STATES
B. SensiaSEDR01
More than 10 numbers, contact manufacturer
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis, Minnesota
55432
UNITED STATES
C. AttestaATDRL1
More than 10 numbers, contact manufacturer
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis, Minnesota
55432
UNITED STATES
A. AdaptaADDR01

Introduction

Affected Products
  1. Adapta
  2. Sensia
  3. Attesta

Reason

Devices in the affected subset, when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality. For this error to occur, a unique combination of events must take place while the device is processing an atrial-sensed event. If this error occurs, the device will be unable to provide pacing until a ventricular-sensed event (VS) is detected. Once a VS is detected, normal pacing functionality is restored immediately. If a VS is not detected, the device will withhold both atrial and ventricular pacing. In addition, until a VS is detected, the device will be unable to initiate a session with a programmer, initiate a session with a CareLink Remote Monitor, or respond to a magnet. Single chamber and dual chamber pacing modes that do not sense atrial activity are not susceptible to this circuit error.